CO2 GAS ANAKLYSER APPARATUS 1
CO2 GAS ANAKLYSER APPARATUS 1

CO2 GAS ANAKLYSER APPARATUS

    oops...!

    no product data.

    Go to homepage

    Application

    This machine used to measure the residual carbon dioxide (dead zone) in the respirator, escape device and other protective masks; it is suitable for respiratory protection product testing institutions and respirator manufacturers to test and evaluate the performance of the masks used by respirator products.

    Features

    1. The core carbon dioxide detection device of this machine is a device that is controlled by a computer program to detect the residual amount of carbon dioxide (dead zone) in the respirator and escape device mask under the state of simulating human breathing.

    2. The test process is controlled by a computer program to realize automatic detection, automatically record the changes in the concentration of inhaled gas carbon dioxide during the entire test process, and display the change curve on the screen.

    3. All the data of the test process are automatically saved in the computer, and the test report is automatically generated. The data of the entire test process can be viewed through the computer at any time, and the test result report can be printed at any time.

    4. The one-way valve that controls the breathing channel is automatically controlled by the ventilator using a solenoid valve group, which is completely synchronized with the ventilator. Make the test results more accurate.

    Standard

    EN149-2001

    Technical Parameters

    1 ventilator

    1.1   Flow characteristics

    Sine   wave

    1.2   Respiration rate

    25   times/min

    1.3   Tidal volume

    2.0L/time

    1.4   Accuracy of breathing volume

    ±5%

    1.5   Accuracy of respiratory rate

    ±5%

    2.   Carbon dioxide analyzer

    2.1   Measuring range

    012%   (three channels)

    2.2   Resolution

    0.1%   (full scale)

    2.3   Linear error

    ≤±2%F.S

    2.4   Repeatability

    ≤±2%F.S

    2.5   Response time constant

    <20   seconds

    3.   Rotameter

    3.1   Range

    0.7-7L/min

    3.2   Accuracy

    2.5   level

    4.   The dead space volume of the gas channel

    not   exceed 2000 ml (except for ventilators)

    5.   Equipped with 3 medium-sized head molds of large, medium and small.

    6. Power Single-phase 220V/50Hz


    Leave A Message

    You can also contact me via email. My email address is manager@qc-test.com
    Related Products
    no data

    

    CONTACT PERSON :Mrs. Zhong
    TEL : +86-0769-8905 5588

    E-MAIL : manager@qc-test.com
    PHONE : +86-13602361535
    FACTORY ADD: No.13 Daxin Street, Daluosha Industrial Zone, Daojiao Town, Dongguan City, Guangdong Province.

    HAIDA Test Equipment

    ISTA Series Test Equipment
    Battery Test Equipment
    Environmental Test Chamber
    Tensile Test Equipment
    Paper Packaging Test Equipment
    Furniture Test Equipment
    Vibration Shaker Table Systems
    Textile Test Equipment
    Mask Making Machine
    Leather & Shoes Test Machine
    Luggage Test Equipment
    Optical Test Equipment
    Rubber and Plastic Test Equipment
    RoHS Testing Equipment
    Cookware Testing Equipment
    Car parts testing equipment

    Contact us

    Copyright © 2024 Haida International Equipment Co., Ltd. | Sitemap
    Customer service
    detect