Temperature Humidity Chamber With PLC Controller
Applications:
Temperature Humidity Chamber It is mainly used to simulate the harsh high temperature, low temperature or damp heat environment that may be encountered during transportation, storage and use of electronic instruments, new materials, electricians, vehicles, metals, electronic products, aerospace materials, etc.; or to simulate high temperature, low temperature, damp heatl In an alternating environment, inspect the high temperature, cold, damp and heat resistance of material accessoriesl or instruments and equipment, as well as the possible damage life reduction, etc.
Electronics
Electronics
Vehicles
Metal Materials
Test standard(including but not limited to... ):
Compliance with ASTM G53-77 specifications GB, GJB, MIL, ASTM, CNS, IEC, JIS, AATCC, DIN, Fed Test, BS and SAE.
1. Internal dimension (W*D*H) ——400*500*500mm
2. External dimension (W*D*H)—— 900*1070*1400mm
3. Temperature range ——-70~+150℃
4. Temperature fluctuation ——±0.5℃
5. Temperature uniformity ——2℃
6. Humidity range ——20~98%RH (Refer Image below)
7. Humidity fluctuation—— ±2.5%RH
8. Humidity uniformity—— 3%RH
9. Cooling speed—— 1℃/min in average (without loading)
10. Heating speed ——3℃/min in average (without loading)
11. Internal chamber material—— SUS#304 stainless steel, mirror finished
12. External chamber material—— Stainless steel
13. Cooling method—— Air cooling
14. Controller LCD touch screen, programmable control temperature and humidity
Can set different parameter for cyclic test
15. Compressor—— France Tecumseh compressor x 2sets
16. Lighting ——Heat resistance
17. Observation window——Tempered glass
Advantage:
1. Temperature range: -70~+150℃
2. Humidity range: 20~98%RH (Refer Image )
3. Cooling speed: 1℃/min in average (without loading)
4. Heating speed: 3℃/min in average (without loading)
5. Internal chamber material: SUS#304 stainless steel, mirror finished
6. High performance and quiet operation (68 dBA)
Application Details:
1. To simulate test environment with different temperature and humidity
2. Cyclic test includes climatic conditions: temperature and humidity test,holding test, cooling-off test, heating-up test, moistening test and drying test…
3. Cable port with flexible silicone plug for cable routing to provide the condition of test unit under operation
4. Uncover test unit weakness in a short-term test with accelerated time effect
In order to control temperature and humidity more accurately, our engineers use intelligent programming controller for the climate chamber, on the one hand can better control the temperature and humidity changes in speed and degree of accuracy. On the other hand, you can simplify the procedure. As long as you set the temperature value, humidity value and test time of the test, the climate chamber can intelligently complete the test and multiple cycle testing. The Intelligent programming controller can display data change trend chart, maximum value, lowest value and so on, and can implement multiple detection data contrast. Automatically saves data and prints test results.
1. PLC controller for test chamber
2. Step types include: ramp, soak, jump, auto-start, and end
3. RS-232 interface to connect computer for output
Model | HD-E702-100 | HD-E702-150 | HD-E702-225 | HD-E702-408 | HD-E702-1000 |
Internal dimension | 400*500*500mm | 500*500*600mm | 500*600*750mm | 600*800*850mm | 1000*1000*1000mm |
External dimension | 900*1070*1400mm | 1000*1070*1500mm | 1000*1170*1650mm | 1100*1370*1750mm | 1500*1380*1900mm |
Voltage(V) | Sigle Phase 220V | Sigle Phase 220V | Sigle Phase 220V | Three Phase 380V | Three Phase 380V |
Temperature range | -20℃,-40℃,-70℃~150℃ |
Temp.& Humi.Adjust Way | Balanced Temperature Humidity Control System(BTHC),Intelligent PID Regulation |
Temperature fluctuations | ≤±0.7℃ (without load and temperature stable) |
Internal Chamber Material | Stainless Steel 304 |
External Chamber Material | Stainless steel or paint spray |
Temperature uniformity | ≤±0.7℃ (without load and temperature stable) |
Temperature deviation | ≤±1.0℃/ ±2.0℃ (without load and temperature stable) |
Heating rate | 3℃/min in average |
Cooling rate | 1℃/min in average |
Humidity range | 20%~98%RH |
Humidity deviation | 1 、≥75%RH:≤±3%RH ;2 、≤75%RH :≤±7%RH |
Cooling Method | Single-stage compression,Second-stage Compression; |
Refrigerator | Hermertically Sealed France Tecumseh Compressor or Semi-hermetic BOCK Compressor |
Cooling Method | Air-cooled |
Door Handle
Adopt non-reaction, explosion-proof door handle,operating easily ,safe and reliable
View Window
Double-layer vacuum-strengthened glass, inner glued sheet conductive film is heated and defrosted. observe clearly the specimen testing situation in the chamber
Test Hole
Φ50mm test hole with thermal Insulation function
Control Pannel
Adopt high precision LCD programmable temperature controller