Corrosion Resistance Tester
Applications:
Corrosion Resistance Tester adapted to the surface treatment of various materials, including paint, electroplating, organic and inorganic coating, anode processing, antirust oil and so on, after these anti-corrosive treatment to test the corrosion resistance of their products.suitable for parts, electronic components, metal materials, protective coatings and industrial products.
Hardware
Electrical Appliances
Automotive
Electronics
Experiment method:
ASTM B117-08, ISO 9227
And IEC, MIL, DIN, ASTM and other standards equivalent to
1, Durable hard 8 mm PVC construction, anti-corrosion and easy to clean.
2, Reservoirs with mixing system on stand.
3, Separate heating system apart from salt testing chamber to ensure accurate result and protect circuit.
4, Exposure zone lined with corrosion-resistant PVC sheets.
5, Clear cover allows viewing of test conditions.
6, Peaked cover prevents dripping of condensation onto samples.
7, Water seal and water level protection system.
8, Heated humidifying tower with air relief valve, adjustable tower tip to orient salt fog.
9, Automatic DI water fills to humidifying tower.
10, Digital temperature control.
11, Integral over-protection on heaters.
12, Low water cut-off protection on humidifying tower.
Method of operation:
1, the salt spray chamber internal parts: tower spray device, funnel, rod, V-type sample holder installed.
2, connected to the gas supply, the power is turned on, turn on the power switch.
3, in the pressure tank and cabinets add water, two low level light exterminate stop adding water, at the water seal groove seal.
4, purified water and sodium chloride salt in a ratio of 1:20 with a good. The solution was poured into the reagent with a good entrance (low saline to exterminate a lamp).
5, we can open all the lamps burn out operation, spray, time switch, switch demisting before opening end of the test.
6, according to the test requirements for equipment and cycles for continuous spraying mode (continuous spray T1, T2 tables are all set to 0, T1 is set to intermittent spray working time, T2 is time to rest).
Model | HD-E808-60 |
Internal Size (W*D*H) | 600*450*400mm |
External Size (W*D*H) | 1060*580*1050mm |
Chamber material | P.V.C rigid plastic plate, thickness 8mm |
Sample tray | Diameter 10mm glass fiber rod, anti-rust V-shape plastic steel to make sure sample in 15~30 degree inclined |
Temperature Range | Room temperature ~ 55℃ |
Temperature Fluctuations | ≤±0.5℃ |
Temperature Uniformity | ≤±2℃ |
Temperature Precision | ±1℃ |
Spraying method | Continuous and Periodic spray type |
Test Chamber Temperature | Salt Spray Method (NSS ACSS)35±1℃ Corrosion-resistant Testing Method (CASS) 50±1℃ |
Saturated air Barrel Temperature | Salt Spray Method (NSS ACSS)47±1℃ Corrosion-resistant Testing Method (CASS) 63±1℃ |
Brine Temperature | 35℃±1℃ |
Spray Quantity | 1.0~2.0 ml / 80cm2 / hr |
PH value | Salt Spray Method (NSS ACSS6.5~7.2) Corrosion-resistant Testing Method(CASS)3.0~3.2 |
Lab Volume | 108L |
Brine Tank Capacity | 25L |
Multiple Safety Protection Devices | Current discharge protection, over pressure protection, over temperature protection, over load fuse protection |
Accessories | Testing Salt/ Measuring tank/Nozzle Equipment x 1 set |
Air source | 1HP Air pump (provided by buyer) |
Presssure Regulating Valve
The pressure regulating valve is used for adjusting the pressure inside the chamber when the air pressure is over high or low
Pressure Gage
The pressure gage can display the pressure value when the heated air reaches the nozzle by going throug the Saturated air tank.
Measuring Pail
Gather the volume of spray of each test 50ml per degree